4 product recalls were published during this week. Review the summaries below, then follow the official source before taking action.

Pharmacy and medical products

Flocare Infinity Enteral Feeding Pumps

Nutricia Australia has received overseas reports of the Flocare Infinity enteral feeding pump 'air in line' (AIL) alarm not always working when used with mixed or reconstituted powdered tube feeds. It is believed that undissolved particles can accumulate in the feeding line or form a film on the inner side of the tube around the pump sensors, which can interfere with the pump's air detection function.

Read summary →
Pharmacy and medical products

Astral 100 and Astral 150 ventilators

ResMed has received a number of reports regarding the performance of the internal battery in Astral 100 and Astral 150 ventilators. According to these reports, an internal electrical issue has led to ventilation stopping without either the 'low battery' or 'critically low battery' alarms activating as intended.