Recall archive/Weekly round-ups/9–15 December 2019
AUSTRALIAN RECALL ROUND-UP

9–15 December 2019.

9 product recalls were published during this week. Review the summaries below, then follow the official source before taking action.

Pharmacy and medical products

Stellar 100 and Stellar 150 Portable Ventilators

The audible alarms may not operate properly. Alarms will display on the screen of the Stellar device. Due to a Q64 electronic component failure on the Stellar device, the identified issue is a combination of software and a component failure. The alarm buzzer may not work under the following combined circumstances: 1. The device has a failed Q64 electronic component and, 2. The device is stored without AC power connected for more than 36 hours leading to full depletion of the battery; and, 3. The device powers on automatically when connected to AC power without pressing the power switch.

Read summary →
Vehicle toys

Emergency Cars 3-Pack

The products do not meet the requirements of the mandatory standard 'Consumer Protection Notice No. 14 of 2003 - Consumer Product Safety Standard: Toys for children up to and including 36 months of age'. The products may break, releasing small parts.

Read summary →
Diving

Body Glove Adult Focus Snorkel 2.0

The snorkel may become blocked by the silicone washer designed to prevent water flowing into the snorkel when submerged. This may result in airflow being unexpectedly restricted and prevent the user from breathing normally when underwater.

Read summary →