T34 Syringe Pump
Published 10 March 2021
Product
T34 Syringe Pump All serial numbers ARTG 131232 (REM Systems Pty Ltd - Infusion pump, syringe) For the provision of small dose infusion therapy, the T34 Syringe Pump gradually administers small precise amounts of fluid and/or medications to a patient.
Why it was recalled
The voltage of the 9V/6LR61 battery may drop prematurely or earlier than expected in T34 2nd and 3rd edition devices. This may cause an interruption to the infusion, following events of: 1. Premature 'low battery' or 'end battery' alarms 2. Unexpected pump shut down with or without alarm. Battery brands/lots that have lower and more stable internal resistance are not likely to cause significant voltage dips and therefore not likely to trigger early low/end battery alarms or pump shut down.
Hazard
If there is an interruption to the infusion, or the pump shuts down unexpectedly, this could result in the patient receiving insufficient delivery of medication or fluids, which can lead to serious health complications.
Source: Based on ACCC data or other content. ACCC © Commonwealth of Australia. ACCC-owned material is licensed under CC BY 4.0. RecallMatch extracts and formats notice information for screening. Third-party material may have separate rights. Source and reuse terms. Verify details in the official notice. RecallMatch is independent of the ACCC.