MyWay Walker Chassis

Published 28 March 2021

Product

MyWay Walker Chassis Product code: 173-1600, 173-2600 and 173-3600 Serial numbers range from LEC/220568 – LEC/272111 ARTG 298435 (Sunrise Medical Pty Ltd - Holder, patient)

Why it was recalled

The rear legs of the walker have a small cylindrical mounting feature called a boss, which the castors are screwed into. If the welding is not sufficient between the leg and the boss, the boss can come loose and the castors will detach from the rear legs of the chassis.

Hazard

If the weld fails, the boss comes loose and the castors detach from the legs of the walker, this could result in an injury and/or a fall.