AUSTRALIAN PRODUCT RECALL

MiniMed 640G and 670G Insulin Pumps

Published 3 November 2021

Product

MiniMed 640G insulin pump An ambulatory, battery-operated, rate-programmable microinfusion pump. Model numbers: MMT-1711 and MMT-1712 MiniMed 670G insulin pump An insulin pump intended for continuous delivery and delivery of user activated insulin boluses. Model numbers: MMT-1762 and MMT-1782 Only pumps with a clear pump retainer ring ARTG 95763 (Medtronic Australasia Pty Ltd Infusion pump, insulin, ambulatory) ARTG 308140 (Medtronic Australasia Pty Ltd - Infusion pump, insulin, ambulatory)

Why it was recalled

If the pump retainer ring breaks, the reservoir may become loose and may not lock into the pump.

Hazard

If the retainer ring is damaged, loose or missing, this may lead to under delivery of insulin, which may cause hyperglycemia. If a user inserts the reservoir back in while the infusion set is still connected to the body, there may be a rapid infusion of insulin, which may cause hypoglycemia or death.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.