AUSTRALIAN PRODUCT RECALL

Medtronic MiniMed Infusion Sets including MiniMed Quick-set, MiniMed mio, MiniMed Silhouette and MiniMed Sure-T (specific lots)

Published 19 September 2017

Product

Medtronic MiniMed infusion sets are used in conjunction with an insulin pump to deliver a continuous dose of insulin to help people with diabetes maintain normal blood glucose levels.

Why it was recalled

There have been recent reports of potential over-delivery of insulin shortly after an infusion set change. Subsequent investigations have found that this issue may be caused by fluid blocking the infusion set membrane during the priming/fill-tubing process. A membrane blocked by fluid most likely occurs if insulin, alcohol or water is spilled on the top of the reservoir or inside the tubing connector, which could prevent the insulin set from working properly.

Hazard

Over-delivery of insulin can cause hypoglycaemia (low blood glucose level), which can lead to serious health complications.

Read the official notice ↗

Source: Based on ACCC data or other content. ACCC © Commonwealth of Australia. ACCC-owned material is licensed under CC BY 4.0. RecallMatch extracts and formats notice information for screening. Third-party material may have separate rights. Source and reuse terms. Verify details in the official notice. RecallMatch is independent of the ACCC.