AUSTRALIAN PRODUCT RECALL

LIFEPAK CR2 Defibrillator

Published 22 April 2021

Product

LIFEPAK CR2 defibrillators are automated external defibrillators (AEDs) used in cases of life-threatening irregular heartbeat that leads to Sudden Cardiac Arrest (SCA) ARTG 282772 (Stryker Australia Pty Ltd - Non-rechargeable public automated external defibrillator)

Why it was recalled

The LIFEPAK CR2's lid magnet may dislodge from the device. The lid magnet is the primary means by which the device will turn on and off when the lid is opened or closed. If the battery dislodges: 1. The battery may be depleted prematurely, even if the device is not powered on; and 2. Users may not be able to turn the device on if they do not use the on/off button or if the battery has fully depleted.

Hazard

If the battery has been depleted or users do not turn the device on, the device will not work and will not be available to defibrillate a patient in an emergency.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.