AUSTRALIAN PRODUCT RECALL

Invacare Perfecto2 V Oxygen Concentrator

Published 19 January 2022

Product

Invacare Perfecto2 V Oxygen Concentrator Product Model: IRC5PO2VAWL2 Serial numbers: 18BF007341 to 181F024339 ARTG: 162637 (Invacare Australia Pty Ltd - Oxygen concentrators)

Why it was recalled

The sound abatement washer may break down resulting in metal on metal wear inside the Pressure Equalisation (PE) valve. This may lead to high temperatures and overheating, separation of the shroud from the device, and loud noise, and can result in a fire that will shut down the device.

Hazard

If the device shuts down this can cause hypoxemia (low oxygen in the blood) from "loss of use". There is an increased risk of fire, which may result in serious injury and property damage.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.