AUSTRALIAN PRODUCT RECALL

Inogen One G3 Concentrator

Published 27 February 2020

Product

Inogen One G3 Concentrator Portable lightweight medical oxygen machine Part Numbers: 19252750, 19252534, 19285707, 19285618, 19285718, 19285621, 19285334, 19252739, 19285727, 19284756, 19252747, 19285626, 19285728, 19252726, 19284685, 19282715, 19285619, 19252733 ARTG 225950 (Air Liquide Healthcare Pty Ltd - Stationary oxygen concentrator)

Why it was recalled

When connected to a fully charged 8-cell battery as its only power source, an operating condition may cause an abrupt shut down of the Inogen One G3 Portable Oxygen Concentrator (POC). If no action is taken, the POC may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with a standard flash screen on display. The POC will then be ready for another startup. If powered on again, this cycle may need to be repeated until the 8-cell battery has been adequately discharged allowing the POC to operate as normal.

Hazard

Risk of interruption to the supply of oxygen to users, which may lead to health complications.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.