AUSTRALIAN PRODUCT RECALL

Fluke-28IIEX

Published 2 April 2013

Product

Intrinsically safe digital multimeter Identifying features description Serial number range 20960000 to 22489999

Why it was recalled

Over time the plastic instrument case may develop cracks near the screws on the back of the unit.

Hazard

If cracks develop and if used in an explosive dust environment (IEC/EU: Zone 21, and US/CA: Class II, Division 1), ingress of conductive dust may cause a short circuit that could ignite an explosive atmosphere surrounding the unit, potentially resulting in serious injury or death.

Read the official notice ↗

Source: Based on ACCC data or other content. ACCC © Commonwealth of Australia. ACCC-owned material is licensed under CC BY 4.0. RecallMatch extracts and formats notice information for screening. Third-party material may have separate rights. Source and reuse terms. Verify details in the official notice. RecallMatch is independent of the ACCC.