AUSTRALIAN PRODUCT RECALL

Flocare Infinity II Enteral Feeding Pumps

Published 15 April 2021

Product

Flocare Infinity II Enteral Feeding Pumps Product codes: 40407 and 40408 All serial numbers ARTG 119667 (Nutricia Australia Pty Ltd - Infusion pump, enteral feeding)

Why it was recalled

In a situation where the device is not plugged into the main power supply and when running on battery power only, the device may unexpectedly switch off without sounding the alarm to indicate a low battery charge. This issue only affects Flocare Infinity Enteral Feeding Pumps. It does not affect any other Nutricia products.

Hazard

If the device switches off, this could cause a delay in the feeding routine and potentially lead to dehydration and diabetic complications.

Read the official notice ↗