Flocare Infinity II Enteral Feeding Pumps
Published 15 April 2021
Product
Flocare Infinity II Enteral Feeding Pumps Product codes: 40407 and 40408 All serial numbers ARTG 119667 (Nutricia Australia Pty Ltd - Infusion pump, enteral feeding)
Why it was recalled
In a situation where the device is not plugged into the main power supply and when running on battery power only, the device may unexpectedly switch off without sounding the alarm to indicate a low battery charge. This issue only affects Flocare Infinity Enteral Feeding Pumps. It does not affect any other Nutricia products.
Hazard
If the device switches off, this could cause a delay in the feeding routine and potentially lead to dehydration and diabetic complications.
Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.