AUSTRALIAN PRODUCT RECALL

Bock Nursing Care Beds - Lifting Poles

Published 12 February 2020

Product

Bock Nursing Care Beds - Lifting Poles Model Number: 423 Serial Number: 91905420-028 ARTG 209294 (Alphacare Pty Ltd - Bed, general-purpose)

Why it was recalled

Beds with attached lifting poles, that were manufactured between April and August 2019, may have faults in the lifting poles.

Hazard

Faulty-shaped lifting pole sockets can lead to slip-through of the lifting pole and may fail to sufficiently support the weight of users, resulting in a risk of injury.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.