AUSTRALIAN PRODUCT RECALL

ABS6090 AEROBURN PE Burn Sheets, 60 cm x 90 cm

Published 4 May 2022

Product

ABS6090 AEROBURN PE Burn Sheets, 60 cm x 90 cm All lots supplied between September 2018 and March 2022 Batch numbers: 186090, 196090, BS690323, BS690916, BS690D14, BS690N08, BS690O10, BS691302, BS691321, BS691625, BS691720 and BS691810 Product code: ABS6090 ARTG 218614 (Aero Healthcare - Burn wrap)

Why it was recalled

The seals on some of the product may become damaged or broken when handled. If the seal on the pouch opens before use, the dressing may lose its microbiological barrier and become non-sterile.

Hazard

If the dressing is non-sterile and used to treat wounds, this may result in infection and health complications.

Read the official notice ↗

Source: Based on ACCC data or other content. ACCC © Commonwealth of Australia. ACCC-owned material is licensed under CC BY 4.0. RecallMatch extracts and formats notice information for screening. Third-party material may have separate rights. Source and reuse terms. Verify details in the official notice. RecallMatch is independent of the ACCC.