AUSTRALIAN PRODUCT RECALL

700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA

Published 31 July 2024

Product

High-pressure pneumatic calibration manifold units included in 700HPPK test kits. Affected models: 700HPPK-NPT – Fluke No. 4784838 700HPPK-BSP – Fluke No. 4810462 700HPPK-MET – 4810443 All units with serial numbers below 6516999 are affected by this recall. Units with serial number 6517000 and above are not affected by this recall.

Why it was recalled

The plunger nut threads on the calibration manifold may fail, causing the plunger assembly to be ejected from the manifold at high speed.

Hazard

There is a risk of serious injury to the user if they are struck by the ejected plunger assembly.

Read the official notice ↗

Source: Product Safety Australia. RecallMatch AU provides a screening summary and is independent of the ACCC.